Medical Device Cybersecurity: FDA 524B, Testing, and Hospital Controls

Medical device cybersecurity is the set of design, testing, and maintenance controls that keep a connected device, and the systems it talks to, from being used to harm patients or expose their data. In the United States it is a statutory requirement. Since March 29, 2023, section 524B of the FD&C Act has required every premarket submission for a cyber device to include a plan for postmarket vulnerabilities, processes that give reasonable assurance the device is cybersecure, and a software bill of materials (SBOM). FDA's premarket cybersecurity guidance, revised in February 2026, lists the evidence reviewers expect: a threat model, a security risk assessment, architecture views, penetration test results, and security labeling. Hospitals own inventory, segmentation, and patching.
This article reflects the rules in force on November 22, 2026. It is not regulatory or legal advice.
Section 524B requirements for cyber devices
Section 524B, codified at 21 USC 360n-2, was added by section 3305 of the Food and Drug Omnibus Reform Act (FDORA), enacted December 29, 2022, and took effect 90 days later. A cyber device meets three tests: it includes software the sponsor validated, installed, or authorized; it has the ability to connect to the internet; and it has technological characteristics that could be vulnerable to cybersecurity threats. FDA reads the middle test broadly. Its cybersecurity FAQ counts Wi-Fi, cellular, Bluetooth, other radio links, and hardware connectors such as USB, Ethernet, and serial ports, whether the connection is intended or not. Under that reading, a device with a USB service port can meet the connectivity test even if it never joins a hospital network.
The sponsor of a 510(k), PMA, PDP, De Novo, or HDE submission for a cyber device must meet four requirements:
- Submit a plan to monitor, identify, and address postmarket cybersecurity vulnerabilities and exploits in a reasonable time, including coordinated vulnerability disclosure.
- Maintain processes that give reasonable assurance the device and related systems are cybersecure, and release patches on a reasonably justified regular cycle for known unacceptable vulnerabilities and as soon as possible out of cycle for critical vulnerabilities that could cause uncontrolled risks.
- Provide an SBOM that covers commercial, open-source, and off-the-shelf software components.
- Comply with any other requirements FDA sets by regulation.
A device submitted before March 29, 2023 sits outside 524B until the manufacturer makes a change that needs a new premarket submission, and that submission falls under the law. FDA's refuse-to-accept transition policy for cyber devices expired on October 1, 2023, and an eSTAR goes on Technical Screening hold if its cybersecurity section lacks accurate responses and relevant attachments. Failing to comply with 524B(b)(2) is a prohibited act under section 301(q)(3) of the FD&C Act, codified at 21 USC 331(q)(3).
What FDA expects in a premarket cybersecurity submission
The working document is Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions, a final guidance issued in February 2026 under docket FDA-2021-D-1158. It supersedes the June 27, 2025 version and aligns the text with the Quality Management System Regulation (QMSR), which took effect on February 2, 2026 and incorporates ISO 13485:2016 by reference. FDA presents a secure product development framework (SPDF) as one way to meet the QMSR for cybersecurity.
Threat modeling comes first. FDA asks manufacturers to model the whole medical device system, to state assumptions such as "hospital networks are inherently hostile," and to cover supply chain, manufacturing, maintenance, and decommissioning. The security risk assessment then scores risk by exploitability instead of probability, because FDA says incident likelihood cannot be reliably estimated from historical data or modeling, and it shows scores before and after mitigation. Any vulnerability in CISA's Known Exploited Vulnerabilities (KEV) Catalog should be designed out of the device.
The architecture section shows controls in eight categories, from authentication and cryptography to event detection and logging and software updates, and it includes four views at minimum: a global system view of every connection, a multi-patient harm view, an updateability and patchability view of the update path, and security use case views covering operational states and clinical functions such as alarming and therapy delivery. FDA's example of multi-patient harm is a fleet of bedside monitors all restarting at once.
FDA recommends a machine-readable SBOM with the NTIA minimum elements, plus each component's level of support (actively maintained, no longer maintained, or abandoned) and its end-of-support date. Each known vulnerability in those components needs a risk assessment with its controls described. Our guide to SBOM requirements covers the same requirement under the EU Cyber Resilience Act and in federal contracts. For labeling, FDA recommends a list of network ports and interfaces, the SBOM in machine-readable form, the procedure for getting authorized updates, backup and restore steps, and the end-of-support date.
Where penetration testing fits in an FDA submission
FDA expects a test package that traces back to the threat model. The guidance lists security requirements testing, threat mitigation testing, and vulnerability testing in line with ANSI/ISA 62443-4-1, which includes abuse and misuse cases, fuzz testing, attack surface analysis, vulnerability chaining, software composition analysis of binaries, and static and dynamic code analysis that checks for hardcoded and default credentials. Penetration testing is the fourth type on the list.
FDA names five elements a penetration test report should cover: the independence and technical expertise of the testers, the scope of testing, the duration, the methods used, and the results, findings, and observations. The report should say who did the testing and how independent they are from the device's designers, and FDA notes that a third party may be necessary for that independence. Third-party reports go in their original form. Any finding deferred to a later release needs a plan with a timeline for reaching fielded devices. After release, FDA recommends testing at regular intervals that match the risk, with annual testing as its example.
Scope deserves the most attention. A test limited to device firmware misses the update server, the cloud service, and the clinician application that appear in the global system view, and those shared components are where one compromise reaches many patients. Build the test plan from the architecture views so each connection has matching test evidence.
Postmarket patching and vulnerability management
Section 524B turns postmarket security into a premarket deliverable. The guidance asks the cybersecurity management plan to name responsible personnel, the sources used to find vulnerabilities such as researchers and the NIST National Vulnerability Database, the handling of KEV entries, periodic security testing, the timeline to release patches, and the coordinated disclosure process. FDA also recommends tracking three measures: the percentage of identified vulnerabilities that are patched, the time from identification to patch, and the time from patch availability to field deployment.
The postmarket cybersecurity guidance issued December 28, 2016 remains FDA's final guidance on how a manufacturer classifies a new vulnerability. Risk is controlled when the residual risk of patient harm is acceptable and uncontrolled when it is not. A routine patch for a controlled risk is generally a device enhancement that does not need to be reported under 21 CFR Part 806. An uncontrolled risk is reportable under Part 806 unless four conditions all hold: no known serious adverse events or deaths are tied to the vulnerability; the manufacturer tells customers and identifies interim compensating controls within 30 days of learning of it; it validates and distributes a fix within 60 days; and it actively participates in an information sharing and analysis organization (ISAO). The FAQ adds that a change made solely to strengthen cybersecurity likely does not need a new premarket submission.
What are the medical device cybersecurity standards?
No single standard is mandatory in the United States. Sponsors cite standards from FDA's recognized consensus standards database, including these.
| Standard | What it covers | FDA recognition number |
|---|---|---|
| ANSI/AAMI SW96:2023 | Security risk management for device manufacturers | 13-131 |
| AAMI TIR97:2019 | Postmarket security risk management for manufacturers | 13-112 |
| IEC 81001-5-1:2021 | Security activities across the health software life cycle | 13-122 |
| ANSI/ISA 62443-4-1-2018 | Secure product development life cycle requirements | 13-119 |
| ANSI/UL 2900-1 and 2900-2-1 | Testable security requirements for network-connectable products, with a healthcare part | 13-96 and 13-104 |
| ANSI/NEMA HN 1-2019 | Manufacturer Disclosure Statement for Medical Device Security (MDS2) | 13-123 |
| IEC 80001-1:2010 | Risk management for IT networks that include medical devices | 13-38 |
SW96 covers risk management, IEC 81001-5-1 or 62443-4-1 covers the development life cycle, and the FDA guidance sets what evidence from each goes into the submission. For EU market access, the Medical Device Regulation places cybersecurity in Annex I. Section 17.2 requires software to be developed and manufactured to the state of the art, taking into account the development life cycle and risk management including information security. Section 17.4 requires manufacturers to set minimum requirements for hardware, IT network characteristics, and IT security measures, including protection against unauthorized access. MDCG 2019-16 is the EU guidance on meeting those sections.
Is there a medical device cybersecurity certification?
FDA does not certify device cybersecurity. FDA clears, grants, or approves the device through the 510(k), De Novo, or PMA pathway, and the cybersecurity documentation is one part of that review. In the EU, where the MDR requires a notified body, that body assesses conformity with Annex I, including sections 17.2 and 17.4, before the manufacturer affixes the CE mark. Voluntary programs also exist: UL Solutions lists UL 2900-2-1 among the standards in its Medical Cybersecurity Assurance Program. A certificate of that kind can support a submission, and the sponsor still has to file the 524B plan, the process evidence, and the SBOM.
Medical device cybersecurity companies fall into four groups: test laboratories that certify products, notified bodies, tool vendors that generate and monitor SBOMs, and consultancies that build documentation and run testing. The firm that writes your threat model should not be the firm that certifies the result. FDA applies the same logic to testing when it asks how independent the testers are from the device designers.
How do hospitals run a medical device cybersecurity risk assessment?
A hospital's assessment covers two harms that FDA treats separately. The 2016 postmarket guidance defines patient harm as physical injury or damage to health, including death, and excludes loss of confidential information such as protected health information, which HIPAA covers. A health system needs a clinical safety view of each device and a HIPAA view of every device that holds electronic protected health information.
HHS gives hospitals a working model in the Health Industry Cybersecurity Practices (HICP), 2023 edition, where practice 9 covers network connected medical devices. For medium organizations it recommends assessing current device security controls, recording inventory traits such as MAC addresses, building security practices for device networks, controlling vendor remote access, and patching after the manufacturer has validated and distributed the patch and it has been tested. For large organizations it adds pre-procurement security requirements, template contract language, and cybersecurity as a stakeholder in clinical procurement.
Inventory holds the program together. FDA's January 30, 2025 safety communication on Contec CMS8000 and Epsimed MN-120 patient monitors is the kind of notice a hospital can act on only if its records show the make, model, and software version of every monitor. At procurement, ask for the MDS2 form, the SBOM, the network port list, and the end-of-support date. Put devices on dedicated segments apart from business systems and route vendor access through one controlled path, as our article on connected operational technology risks describes for building systems. Our healthcare cybersecurity guide covers the wider program.
BD Emerson's penetration testing services cover the parts of a medical device system that face the outside world: web applications, APIs, networks, and cloud back ends, with retesting included. For a manufacturer preparing a premarket submission, we scope the test from your architecture and data flows, and every finding comes with reproduction steps, evidence, severity, and a fix path. For a health system, we test the device networks and vendor access paths that HICP practice 9 asks you to control. FDA makes the clearance decision, and our job is to give you independent test evidence to put in front of it.